Transpulmonary Thermodilution Detects Rapid and Reversible Increases in Lung Water Induced by PEEP in ARDS

link.springer.com
transpulmonary-thermodilution-detects-rapid-and-reversible-increases-in-lung-water-induced-by-positive-end-expiratory-pressure-in-ards

In ARDS patients, changing the positive end-expiratory pressure (PEEP) level induced parallel, small and reversible changes in EVLW. These changes were not due to an artefact of the TPTD technique and were likely due to the PEEP-induced changes in CVP, which is the backward pressure of the lung lymphatic drainage.

Reducing PEEP from HighPEEPstart (14 ± 2 cmH2O) to LowPEEP15′ significantly decreased EVLW from 20 ± 4 to 18 ± 4 mL/kg, central venous pressure (CVP) from 15 ± 4 to 12 ± 4 mmHg, the arterial oxygen tension over inspired oxygen fraction (PaO2/FiO2) ratio from 184 ± 76 to 150 ± 69 mmHg and lung volume by 144 [68–420] mL.

In 60 ARDS patients, we measured EVLW (PiCCO2 device) at a PEEP level set to reach a plateau pressure of 30 cmH2O (HighPEEPstart) and 15 and 45 min after decreasing PEEP to 5 cmH2O (LowPEEP15′ and LowPEEP45′, respectively).